Commands Listed in Alphabetical Order
Examples
EXPORT:VIEW FU
LLSCREEN
sets the exported view area to Full Screen.
EXPORT:VIEW? might return :EXPORT:VIEW GRATICULE, indicating that only
the graticule
area of the screen will be exported.
FACtory (No Query Form)
This command (no query form) resets the instrument to its factory default settings.
This command is equivalent to pressing the DEFAULT SETUP button located on
the instrument front panel or selecting Recall Default Setup from the File menu.
This command resets the instrument settings to factory defaults:
Clears the Event Status Enable Re gister.
Clears t
he Service Request Enable Register.
Sets the Device Event Status Enable Register to 255.
Sets the Power On Status Clear Flag to TRUE.
Purges all defined aliases.
Sets the macro definedby*DDTtoa“zero-lengthfield”.
Clears the pending operation flag and associated operations.
This
command does not reset the following:
The state of the GPIB (IEEE Std 488.1-1987) interface.
The selected GPIB address.
Calibration data that affects device specifications.
Protected user data.
Stored settings.
T
he current password (if you set one).
Group
Save and Recall
Syntax
FACtory
Related Commands
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
Arguments
None
2-216 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual