Commands Listed in Alphabetical Order
Arguments
<NR3> species
the LOW threshold in volts.
Examples
SEARCH:SEARCH1:TRIGGER:A: TRANSITION:THRESHOLD:LOW:CH2 0.5 sets
the Channel 2 transition trigger LOW threshold setting for search 1 to 0.5 volts.
SEARCH:SEARCH1:TRIGGER:A: TRANSITION:THRESHOLD:LOW:CH1? might
return
:SEARCH:SEARCH1:TRIGGER:A:TR ANSITION:THRESHOLD:LOW:CH1
800.0000E–3
, indicating that the Channel 1 transition trigger LOW threshold
setting for search 1 is set to 0.8 volts.
SEARCH:SEARCH<x>:TRIGger:A:TRANsition:THReshold:LOW:MATH<x>
This command sets or queries the threshold low voltage level setting for a
transition trigger search to determine where to place a mark. <x> is the search
number or m ath waveform number.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger: A:TRANsition:THReshold:LOW:MATH<x>
<NR3>
SEARCH:SEARCH<x>:TRIGger: A:TRANsition:THReshold:LOW:MATH<x>?
Arguments
<NR3> species the LOW threshold in volts.
Examples
SEARCH:SEARCH1:TRIGGER:A: TRANSITION:THRESHOLD:LOW:MATH2 0.5
sets the M ath 2 waveform transition trigger LOW threshold setting for search 1
to 0.5 volts.
SEARCH:SEARCH1:TRIGGER:A: TRANSITION:THRESHOLD:LOW:MATH1?
might return
:SEARCH:SEARCH1:TRIGGER:A :TRANSITION:THRESHOLD:LOW:MATH1
800.0000E–3
, indicating that the Math 1 waveform transition trigger LOW
threshold setting for search 1 is set to 0.8 volts.
SEARCH:SEARCH<x>:TRIGger:A:TRANsition:THReshold:LOW:REF<x>
This command sets or queries the threshold low voltage level setting for a
transition trigger search to determine where to place a mark. <x> is the search
number or reference waveform number.
Group
Search and Mark
2-476 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual