Commands Listed in Alphabetical Order
Syntax
*RST
Related Commands
FACtory, RECAll:SETUp, SAVe:SETUp
Arguments
None
Examples
*RST resets the instrument s ettings to factory defaults.
*SAV (No Query Form)
This command (no query form) stores the state of the instrument to a specified
memory location. You can later use the *RCL command to restore the instrument
to this saved state.
This command is e quivalent to s electing Save As from the File menu, and then
pressing the Setup button.
Group
Save and Recall
Syntax
*SAV <N R1>
Related Commands
*RCL, RECAll:SETUp, SAVe:SETUp
Arguments
<NR1> specifies a location in which to save the state of the instrument. Location
values range from 1 through 10. Using an out-of-range location value causes an
execution error. Any settings that have been stored previously at this location
will be overwritten.
Examples
*SAV 2 saves the current instrument state in memory location 2.
SAVe:MASK (No Query Form)
This command (no query form) saves the user mask to a mass storage file.
Group
Mask
Syntax
SAVe:MASK <QString>
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-425