Commands Listed in Alphabetical Order
Examples
TRIGGER:MULTI
SCOPE:ALIGN:SETSTATE Start
causes the instrument to
measure the round trip time between sending a trigger out the TekLink port and
receiving the trigger pulse back from the port.
TRIGGER:MULTISCOPE:ALIGN: SETSTATE? might return
:TRIGGER:MULTISCOPE:ALIGN :SETSTATE DONE, indicating that
the c alibration is complete.
TRIGger:MULTiscope:ALIGN:DESKEW? (Query Only)
This command sets o r queries the deskew time, in nanoseconds, for horizontal
position
ing of the waveform .
Group
TekLink
Syntax
TRIGger:MULTiscope:ALIGN: DESKEW?
Examples
TRIGGER:MULTISCOPE:ALIGN: DESKEW? might return
:TRIGGER:MULTISCOPE:ALIGN :DESKEW 2.50000E–9, indicating
that the deskew time is 2.5 ns.
TRIGger:MULTiscope:ALIGN:MRTTime
This command returns or sends the maximum round trip time to each of the
instruments in a calibration cycle.
Group
TekLink
Syntax
TRIGger:MULTiscope:ALIGN: MRTTime
TRIGger:MULTiscope:ALIGN: MRTTime?
Examples
TRIGGER:MULTISCOPE:ALIGN: MRTTIME? might return
:TRIGGER:MULTISCOPE:ALIGN :MRTTIME 20.000E–6, indicating
that the round trip time is 20 us.
TRIGger:MULTiscope:ALIGN:VALue? (Query Only)
This command returns the time interval result from the
TRIGGER:MULTISCOPE:ALIGN command. Round trip time, trigger
out to trigger in.
2-620 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual