Commands Listed in Alphabetical Order
0.00000000000
;:HORIZONTAL:MAIN:INTERPRATI O
1.0000;POSITION 50.0000;UNITS:STRIN G "s";:HORIZONTAL:ROLL
AUTO;MODE AUTO;MODE:AUTO:LIMITRECORDLEN
200000;:HORIZONTAL:MODE:RE CORDLENGTH 1000;SAMPLERAT E
40.0000E+9;SCAL
.
HORizontal:ACQDURATION? (Query Only)
This query
returns the timebase duration.
Group
Horizontal
Syntax
HORizontal:ACQDURATION?
Returns
<NR3>
Examples
HORIZONTAL:ACQDURATION? might return :HORIZONTAL:ACQD
URATION
5.0E-9
, indicating the acquisition duration is 5.0 us.
HORizontal:ACQLENGTH? (Query Only)
This query returns the rec ord length.
Group
Horizontal
Syntax
HORizontal:ACQLENGTH?
Returns
<NR1>
Examples
HORIZONTAL:ACQLENGTH? might return :HORIZONTAL:ACQLENGTH 50000,
indicating the acquisition record length is 50000 points.
HORizontal:DIVisions? (Query Only)
This query-only command returns the number of graticule divisions over w hich
the waveform is displayed.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-239

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals