Commands Listed in Alphabetical Order
HORIZONTAL:MO
DE:RECORDLENGTH?
might return
:HORIZONTAL:MODEL:RECORDL ENGTH 1000000, indicating that the record
length is set to 1,000,000 samples
HORizontal:
MODE:SAMPLERate
This command sets or queries the sample rate.
Group
Horizontal
Syntax
HORizontal:MODE:SAMPLERat e <NR1>
HORizontal:MODE:SAMPLERat e?
Arguments
<NR1> is the sample rate in samples per se cond.
Examples
HORIZONTAL:MODE:SAMPLERAT E 1e6 s ets the sample rate to 1 million samples
per second.
HORIZONTAL:MODE:SAMPLERAT E? might return
:HORIZONTAL:MODEL:SAMPLER ATE 5.0000E+6, indicating that the sample
rate is set to 5 million samples per second.
HORizontal:MODE:SCAle
This command sets or queries the horizontal scale.
Group
Horizontal
Syntax
HORizontal:MODE:SCAle <NR1>
HORizontal:MODE:SCAle?
Arguments
<NR1> is the horizontal scale in seconds per division. The h orizontal scale is
read only for Manual mode.
Examples
HORIZONTAL:MODE:SCALE 2e-9 sets the horizontal scale to 2 ns per division.
HORIZONTAL:MODE:SCALE? migh
treturn
:HORIZONTAL:MODEL:SCALE
20.0000E-6
, indicating that the horizontal scale is set to 10 μs per division.
2-262 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals