Commands Listed in Alphabetical Order
TRIGGER:B:PUL
SE:WINDOW:POLARITY?
might return
:TRIGGER:B:PULSE:WINDOW:PO LARITY NEGATIVE, indicating that
the window polarity is set to negative.
TRIGger:B:PULse:WINdow:POLarity:CH<x>
This command sets or queries the pulse trigger w indow polarity of the specified
channel. The oscilloscope triggers only when the signal exits the window. This
command is available only when the Window Event is set to Inside > t and not
available for the rest of the window events. The logic selection is available only
when the p olarity is set to Either. <x> is the search number or channel number.
Group
Trigger
Syntax
TRIGger:B:PULse:WINdow:POL arity:CH<x>
{EITher|NEGAtive|POSITIVe}
TRIGger:B:PULse:WINdow:POL arity:CH<x>?
Arguments
EITher specifies positive or negative polarity.
NEGAtive specifies a negative polarity.
POSITIVe specifies a positive polarity.
Examples
TRIGGER:B:PULSE:WINDOW:POL ARITY:CH<x> POSITIVE sets the window
polarity to positive.
TRIGGER:B:PULSE:WINDOW:POL ARITY:CH<x>? might return
:TRIGGER:B:PULSE:WINDOW:PO LARITY:CH<x> NEGATIVE, indicating that
the window polarity is set to negative.
TRIGger:B:RESET:SOUrce
This command sets or queries the trigger source for the A→B sequential trigger
reset feature, except for the Timeout trigger type.
Group
Trigger
Syntax
TRIGger:B:RESET:SOUrce {CH<x>|AUXiliary}
TRIGger:B:RESET:SOUrce?
Related Commands
TRIGger:B:RESET:TYPe, TRIGger:B:RESET:THReshold
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-611