Commands Listed in Alphabetical Order
Examples
:CALIBRATE:PR
OBESTATE:CH1?
might return
:CALIBRATE:PROBESTATE:CH1 0, indicating that the probe calibration for
Channel 1 passed.
CALibrate:R
ESults? (Query Only)
This query-only command returns the status of internal and factory calibrations,
without performing any calibration operations. The results returned do not include
the calibr
ation s tatus of attached probes. The
CALibrate:RESults? query is
intended to support GO/NoGO testing of the instrument calibration readiness: all
returned results should indicate PASS status if the instrument is “fitforduty”. It
is quite common, however, t o use uncalibrated probes (particularly when the
instrument inputs are connected into a test system with coaxial cables).
This command is equivalent to selecting Instrument Calibration from the Utilities
menu and then viewing the contents of the Status field.
Group
Calibration
Syntax
CALibrate:RESults?
Related Commands
*CAL?
Examples
CALIBRATE:RESULTS? returns the status of all the calibration subsystems.
The query might return
:CALIBRATE:RESULTS:SPC PASS;FACTORY PASS,
indicating that the status of internal and factory calibrations, without performing
any calibration operations is set to PASS.
CALibrate:RESults:SPC? (Query Only)
This query-only command returns the results of the last SPC operation. However,
this query does not cause an SPC to be run. This command is equivalent to
selecting Instrument Calibration from the Utilities menu.
Group
Calibration
Syntax
CALibrate:RESults:SPC?
Related Commands
*CAL?
2-106 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual