Commands Listed in Alphabetical Order
Related Commands
TRIGger:A:HOL
Doff:TIMe
Arguments
TIMe enables you to set the holdoff time via the TRIGger:A:HOLDoff:TIMe
command.
DEFAult automatically calculates a holdoff time to use. This time is typically
equivalent to the g reater of 1/2 screen (5 divisions) of time or 250 ns. The
maximum value is 12 s. For example, if the instrument is set to 1 ms/division then
the default holdoff will be 1 ms/division x 25 divisions = 25 ms.
Examples
TRIGGER:A:HOLDOFF:BY TIME sets the holdoff to the “by time” setting. This
enables you to set the holdoff time.
TRIGGER:A:HOLDOFF:BY? might return :TRIGGER:A:HOLDOFF: BY TIME,
indicating that you will set the holdoff time.
TRIGger:A:HOLDoff:TIMe
This command sets or queries the A trigger holdoff time. This command is
equivalent to selecting Holdoff from the Trig menu and then choosing the desired
Trig Holdoff.
Group
Trigger
Syntax
TRIGger:A:HOLDoff:TIMe <NR3>
TRIGger:A:HOLDoff:TIMe?
Related Commands
TRIGger:A:HOLDoff:BY
Arguments
<NR3> specifies the holdoff time in seconds. The range is from 250 ns through
12.0 s.
Examples
TRIGGER:A:HOLDOFF:TIME 10 sets the A trigger holdoff time to 10 s.
TRIGGER:A:HOLDOFF:TIME? might return :TRIGGER:A:HOLDOFFTIM E
1.2000E-06
, indicating that the A trigger holdoff time is set to 1.2 µs.
TRIGger:A:I2C:ADDRess:MODe
DPO7000 Series only: This command sets or queries the I2C address mode to 7
or 10-bit.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-567