Commands Listed in Alphabetical Order
TRIGger:A:SP
I:SS:ACTIVE
DPO7000 Series only: This command sets or queries the SPI trigger Slave Select
(SS) polarity.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:SPI:SS:ACTIVE {HIGH|LOW}
TRIGger:A:SPI:SS:ACTIVE?
Arguments
HIGH species HIGH for SS polarity.
LOW species LOW for SS polarity.
Examples
TRIGGER:A:SPI:SS:ACTIVE HIGH sets the SPI trigger SS polarity to HIGH.
TRIGGER:A:SPI:SS:ACTIVE? might return :TRIGGER:A:SPI:SS:AC TIVE
LOW
, indicating that the SPI trigger SS polarity is set to LOW.
TRIGger:A:SPI:SS:LEVel
DPO7000 Series only: This command sets or queries the threshold for the SPI
trigger Slave Select (SS) signal.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:SPI:SS:LEVel <NR3>
TRIGger:A:SPI:SS:LEVel?
Arguments
<NR3> speci es the threshold for the SPI trigger SS line. The threshold range
is ±5.0.
Examples
TRIGGER:A:SPI:SS:LEVEL 0.0 sets the threshold level for the SPI trigger
SS to 0.0.
TRIGGER:A:SPI:SS:LEVEL? might return :TRIGGER:A:SPI:SS:LEVEL
5.0000
, indicating that the threshold level for the SPI trigger SS is set to 5.0.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-597

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals