Commands Listed in Alphabetical Order
or optical meas
urement defined. Measurements are specified by x, which ranges
from 1 through 8.
Conditions
This command is only valid if the instrument has Option MTM is installed.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:NOISe {HIGH|LOW}
MEASUrement:MEAS<x>:NOISe ?
Arguments
HIGH argument causes the measurement for noise to be taken at the high level
of the waveform.
LOW argument causes the measurement for noise to be taken at the low level of
the waveform.
Examples
MEASUREMENT:MEAS1:NOISE HIGH causes the noise measurement to be made
at the high level of the waveform for Measurement 1.
MEASUREMENT:MEAS2:NOISE? might return : MEASUREMENT:MEAS2:NOISE
LOW
, indicating that Measurement 2 will measure noise at the low level of the
waveform.
MEASUrement:MEAS<x>:REFLevel? (Query Only)
This query-only command returns the current reference level parameters. It returns
them in the following order: ABSOLUTE and then PERCENT for individual user
measurements. Measurements are spe cifi ed by x, which ranges from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:REFLe vel?
Examples
MEASUREMENT:MEAS1:REFLEVE L? might return
:MEASUREMENT:MEAS1:REFLEV EL:ABSOLUTE:HIGH 1.7100E+00, indicating
that the reference level is set to 1.71 V.
2-388 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual