Commands Listed in Alphabetical Order
POSITIVe enabl
es testing on positive pulses. This is the default.
Examples
MASK:POLARITY BOTH sets pa
ss/fail to test both positive (normal) and negative
(inverted) waveforms.
MASK:POLARITY? might ret
urn
:MASK:POLARITY POSITIVE, indicating that
the input waveform polarity for the pass/fail test is set to positive.
MASK:SEG<m> (No Query Form)
This command deletes the specied mask segment from the current mask.
Group
Mask
Syntax
MASK:SEG<m> DELEte
Arguments
<m> is an integer that species the mask segment number to de lete from the
current mask.
DELETE removes the specied mask segme nt from the mask.
Examples
MASK:SEG3 DELETE deletes mask segment three from the current mask.
MASK:SEG<m>:NR_Pt? (Query Only)
This query-only command returns the number of points that make up the specied
mask segment of the current mask. Each mask point consists of an X- Y pair of
coordinates.
Group
Mask
Syntax
MASK:SEG<m>:NR_Pt?
Arguments
<m> is an integer number that s pecies a mask segment number of the current
mask.
Returns
<NR1> is 0 to 50.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-315

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals