Commands Listed in Alphabetical Order
Table 2-15: DAT
a and WFMOutpre Parameter Settings
WFMOutpre Settings
DATa:ENCdg
Setting :ENCdg :BN_Fmt :BYT_Or :BYT_NR
ASCii ASC N/A N/A
1,2,4,8
FAStest
BIN
RI/FP MSB
1,2,4
RIBinary BIN RI
MSB
1,2,8
RPBinary BIN RP
MSB
1,2,8
FPBinary BIN FP
MSB
4
SRIbinary
BIN RI
LSB
1,2,8
SRPbinary
BIN RP
LSB
1,2,8
SFPbinary
BIN FP
LSB
4
Examples
DATA:ENCDG RPBinary sets the data e ncoding format to be a positive integer
where the most signicant byte is transferred rst.
DATA:ENCDG? might return :DATa:ENCDG SRPBINARY for the format of the
outgoing w aveform data .
DATa:FRAMESTARt
This command sets or queries the starting data frame for waveform transfer. This
command allows for transferring partial frames of the waveform to and from
the instrument.
Group
Waveform Transfer
Syntax
DATa:FRAMESTARt <NR1>
DATa:FRAMESTARt?
Arguments
<NR1> is the rst data frame that is transferred, which ranges from 1 to the number
of acquired frames. Data is transferred from <NR1> to DATa:FRAMESTOP or
the number a acquired frames, whichever is less. If <NR1> is greater than the
number of acquired frames, then you will receive the last acquired frame. If no
frames were acquired, then you receive the rst frame which contains all nulls.
Returns
The starting data frame f or waveform transfer.
Examples
DATa:FRAMESTARt 10 species that the waveform transfer wi
ll begin with
frame 10.
2-156 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual

Table of Contents

Other manuals for Tektronix DPO70000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DPO70000 Series and is the answer not in the manual?

Tektronix DPO70000 Series Specifications

General IconGeneral
BrandTektronix
ModelDPO70000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals