Commands Listed in Alphabetical Order
The instrument
will only display the bar if the associated trigger source is also
displayed. If both a main and a delayed trigger are displayed, then two bars will
appear. One will accompany each source. If a logic trigger is sele cted, then
multiple bars may appear. One will show the upper threshold and one will show
the lower threshold. This command is equivalent to selecting Display Setup from
the Display menu and then choosing the Objects tab.
Group
Display Control
Syntax
DISplay:TRIGBar {OFF|SHORt|LONG}
DISplay:TRIGBar?
Arguments
OFF removes the trigger indicator bar from the display.
SHORt displays, as the indicator, a short arrow at the right side of the graticule for
each displayed trigger signal.
LONG displays, as the indicator, a horizontal line across the width of the graticule
for each displayed trigger signal.
Examples
DISPLAY:TRIGBAR LONG sets the display to show a long trigger indicator bar
(or bars).
DISPLAY:TRIGBAR? might return :DISPLAY:TRIGBAR OFF, indicating that the
indicator bar is removed from the display.
DISplay:TRIGT
This command controls or queries the display of the trigger T. The trigger T shows
where the trigger occurred on the waveform.
Group
Display Control
Syntax
DISplay:TRIGT {<NR1>|OFF|ON}
DISplay:TRIGT?
A
rguments
OFF removes the trigger indicator T from the display.
ON displays a T at the trigger point.
<NR1> = 0 disables the trigger T; any other value displays the trigger T.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-199