Commands Listed in Alphabetical Order
Group
Mask
Syntax
MASK:TESt:DELay <NR3>
MASK:TESt:DELay?
Arguments
<NR3> is a floating point number that specifies the amount of time, in seconds,
to delay the start of the pass/fail test.
Examples
MASK:TEST:DELAY 2.5 delays waveform evaluation by 2.5 seconds a fter the
start of pass/fail testing.
MASK:TEST:DELAY? might return :MASK:TE ST:DELAY 5, indicating that the
amount of time to wait after starting pass/fail testing to evaluate waveforms is
set to 5 seconds.
MASK:TESt:HARDCopy
This command sets o r returns the hard copy in pass/fail failure mode. When
enabled, this command causes the instrument to generate a s creen hard copy, to
the default printer using the current instrument hard copy settings, as soon as the
pass/fail status changes to ‘Failing’.
Group
Mask
Syntax
MASK:TESt:HARDCopy {OFF|ON|<NR1>}
MASK:TESt:HARDCopy?
Arguments
ON turns on generate hard copy on failure.
OFF turns off generate hard copy on failure.
<NR1> = 0 turns off generate hard copy on failure, and any other value turns on
generate hard copy on failure.
Examples
MASK:TEST:HARDCOPY ON turns on generate hard copy on failure.
MASK:TEST:HARDCOPY? might return :MASK:TEST:HARDCOPY ON, indicating
that the generate hard copy on failure is turned on.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-327