Commands Listed in Alphabetical Order
Arguments
ACCept specifie
s that the instrument will only trigger on pulses that are narrower
than the specified width, when the trigger type is set to glitch. The width is
specified using the TRIGger:{A|B}:PULse:GLItch:WIDth command.
REJect specifies that the instrument will only trigger on pulses that are wider than
the specified width, when the trigger type is set to glitch. The width is specified
using the TRIGger:{A|B}:PULse:GLItch:WIDth command.
Examples
TRIGGER:A:PULSE:GLITCH:TRI GIF REJECT specifies that the instrument
triggers on pulses that are wider than the specified width.
TRIGGER:A:PULSE:GLITCH:TRI GIF ACCEPT? might return
:TRIGGER:A:PULSE:GLITCH:TR IGIF ACCEPT, indicating that the instrument
is set to trigger on pulses that are narrower than the specified width.
TRIGger:{A|B}:PULse:GLItch:WIDth
This co
mmand sets or queries the width for the glitch trigger. This command
is equivalent to selecting Glitch Setup from the Trig menu and then setting the
desired Width.
For information about using the width value, refer to the TRIGger:{A|B}:PULse:
GLItch:TRIGIF command.
Group
Trigger
Syn
tax
TRIGger:{A|B}:PULse:GLItch :WIDth <NR3>
TRIGger:{A|B}:PULse:GLItch :WIDth?
Related Commands
TRIGger:{A|B}:LEVel, TRIGger:{A|B}:PULse:GLItch:TRIGIF
Arguments
<NR3> argument specifies the width of the glitch in seconds.
Examples
TRIGGER:A:PULSE:GLITCH:WID TH 15E-6 sets the width of the glitch to 15 µs.
TRIGGER:A:PULSE:GLITCH:WID TH? might return
:TRIGGER:A:PULSE:GLITCH:WI DTH 2.0000E-09, indicating that the width
of the glitch is c urrently set at 2 ns.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-525