Commands Listed in Alphabetical Order
RMSNoise measu
res the Root Mean Square noise amplitude on a waveform at
the mid reference level.
SIGMA1 (histo
gram measurement) measures the percentage of points in the
histogram that are within one standard deviation of the histogram mean.
SIGMA2 (his
togram measurement) measures the percentage of points in the
histogram that are w ithin two standard deviations of the histogram mean.
SIGMA3 (his
togram measurement) measures the percentage of points in the
histogram that are within three standard deviations of the histogram mean.
SIXSigmaj
it
(histogram measurement) is six × RMSJitter.
SNRatio measures the signal-to-noise ratio. The signal-to-noise ratio is the
amplitud
eofanoiserejectionbandcenteredonthemidlevel.
STDdev measures the standard deviation (Root Mean Square (RMS) deviation) of
all acqu
ired points within or on the histogram box.
UNDEFINED is the default measurement type, wh ich indicates that no measurement
type is
specified. Once a measurement type is chosen, it can be cleared using
this argument.
WAVEF
ORMS
(waveform count) measures the number of waveforms used to
calculate the histogram.
Examples
MEA
SUREMENT:IMMED:TYPE FREQUENCY
defines the immediate measurement
to be a frequency measurement.
MEA
SUREMENT:IMMED:TYPE?
might return :MEA SUREMENT:IMMED:T YPE RMS,
indicating that the immediate measurement is the true Root Mean Square voltage.
MEASUrement:IMMed:UNIts? (Query Only)
This query-only command returns the units of the immediate measurement.
Group
Measurement
Syntax
MEASUrement:IMMed:UNIts?
Examples
MEASUREMENT:IMMED:UNITS? might return : MEASUREMENT:IMMED:UNIts
"s"
, indicating that units for the immediate measurement are in seconds.
2-382 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual