Commands Listed in Alphabetical Order
MEASUREMENT:M
EAS2:SOURCE1:SIGTYPE?
might return
:MEASUREMENT:MEAS2:SOURCE1 :SIGTYPE EYE, indicating that the
signal type associated with source 1 of measurement slot 2 is EYE.
MEASUrement
:MEAS<x>:STATE
This command sets or queries whether the specified measurement slot is computed
and displayed. The measurement slot is specified by x, which ranges from 1
through 8.
This command is equivalent to selecting Measurement Setup from the
Measure menu and then clicking the Display button.
For a meas
urement to display, you must have selected a source waveform
and defined the mea surement you want to take and display. You select the
measurement using the MEASUrement:MEAS<x>:SOUrce<x> comma nd.
You define the measurement type using the MEASUrement:MEAS<x>:TYPe
command.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:STATE {OFF|ON|<NR1>}
MEASUrement:MEAS<x>:STATE?
Related Commands
MEASUrement:MEAS<x>:SOUrce<x>, MEASUrement:MEAS<x>:TYPe
Arguments
OFF disables calculation and display of the specified measurement slot.
ON enables calculation and display of the specified measurement slot.
<NR1> = 0 disables calculation and display of the specified measurement slot; any
other value enables calculation and display of the specified measurement slot.
Examples
MEASUREMENT:MEAS2:STATE ON computes and displays the measurement
defined as measurement 2.
MEASUREMENT:MEAS1:STATE? might return :MEASUREMENT:MEAS1:STATE 0,
indicating that measurement defined for measurement slot 1 is disabled.
MEASUrement:MEAS<x>:STDdev? (Query Only)
This query-only command returns the standard deviation of values accumulated
for this measurement since the last statistical reset. Measurements are specified
by x, which ranges from 1 through 8.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-395