Commands Listed in Alphabetical Order
TRIGger:A:VI
Deo:CUSTom:LINEPeriod
DPO7000 Series only: This command sets or queries the period horizontal line
scan. Use this command only when the video format is set to Bilevel or Trilevel.
This is equiv
alent to selecting Video Setup from the Trig menu, setting Bilevel
or Trilevel for Format, and setting the Line Period.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:LIN EPeriod <NR3>
TRIGger:A:VIDeo:CUSTom:LIN EPeriod?
Arguments
<NR3> is the time for each horizontal scan line.
Examples
TRIGGER:A:VIDEO:CUSTOM:LIN EPERIOD 63. 6E-6 sets the custom line period
to 63.6 microseconds.
TRIGGER:A:VIDEO:CUSTOM:LIN EPERIOD? might return TRIGGER:A:
VIDEO:CUSTOM:LINEPERIOD
63.6000E-6 indication the line period is set to
63 microseconds.
TRIGger:A:VIDeo:CUSTom:SYNCInterval
DPO7000 Series only: This command sets or queries the horizontal line s can rate
of the A video trigger. Use this command only when the video format is set to
Bilevel. This is equivalent to selecting Video Setup from the Trig menu, setting
Bilevel for Format, and setting the sync interval.
Group
Trigger
Syntax
TRIGger:A:VIDeo:CUSTom:SYN CInterval <NR3>
TRIGger:A:VIDeo:CUSTom:SYN CInterval?
Arguments
<NR3> is the sync interval for the Bilevel f orma
t.
Examples
TRIGGER:A:VIDEO:CUSTOM:SYN CINTERVAL 4.7E-6 sets the sync interval to
4.7 microseconds.
TRIGGER:A:VIDEO:CUSTOM:SYN CINTERVAL? might return TRIGGER:A:
VIDEO:CUSTOM:SYNCINTERVAL 4.7000E-6
,indica
ting that the custom sync
interval is set to 4.7 microseconds.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-601