Commands Listed in Alphabetical Order
TRIGger:{A|B
}:PULse:WINdow:THReshold:LOW
This c o mmand sets or queries the lower limit for the window t rigger. This
command is equivalent to selecting Window Setup from the Trig menu and setting
the Lower Lev
el voltage.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WINdow :THReshold:LOW <NR3>
TRIGger:{A|B}:PULse:WINdow :THReshold:LOW?
Arguments
<NR3> specifies the threshold value in volts.
Examples
TRIGGER:A:PULSE:WINDOW:THR ESHOLD:LOW 0.8 sets the lower threshold of
the window trigger to 0.8 V.
TRIGGER:A:PULSE:WINDOW:THR ESHOLD:LOW? might return
:TRIGGER:A:PULSE:WINDOW:TH RESHOLD:LOW 1.2000E-01, indicating that
the lower limit of the window trigger is currently set to 120 mV.
TRIGger:{A|B}:PULse:WINdow:WIDTH
This command sets or queries the minimum width for a window violation. This
command is equivalent to selecting Window Setup from the Trig menu, selecting
Wider than in the Trigger When box, and setting the Width.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:WINdow :WIDTH <NR3>
TRIGger:{A|B}:PULse:WINdow :WIDTH?
Arguments
<NR3> argument spec ifi es the minimum width in seconds.
Examples
TRIGGER:A:PULSE:WINDOW:WID TH 15E-9 sets the minimum width of a valid
window v
iolation to 15 ns.
TRIGGER:A:PULSE:WINDOW:WID TH? might return
:TRIGG
ER:A:PULSE:WINdow:WIDTH 2.0000E-09
indicating that the
minimum width of a valid window violation is 2 ns.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-549