Commands Listed in Alphabetical Order
Examples
TRIGGER:A:CAN
:DATA:DIRECTION WRITE
sets the CAN data direction to
Write.
TRIGGER:A:CAN:DATA:DIRECTI ON? might return
:TRIGGER:A:CAN:DATA:DIRECT ION READ, indicating that the CAN data
direction is set to Read.
TRIGger:A:CAN:DATa:LEVel
CAN option only: This command sets or queries the CAN Trigger threshold for
the CAN data source.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:CAN:DATa:LEVel <NR3>
TRIGger:A:CAN:DATa:LEVel?
Arguments
<NR3> specifies the C AN trigger data level. The level can be ±5.0.
Examples
TRIGGER:A:CAN:DATA:LEVEL 0.0 sets the CAN trigger data level to 0.0 Volts.
TRIGGER:A:CAN:DATA:LEVEL? might return :TRIGG ER:A:CAN:DATA:LEVEL
5.0000
, indicating that the CAN trigger data level i
s set to 5.0 volts.
TRIGger:A:CAN:DATa:SOUrce
CAN option only: This command sets or queries the CAN data source.
Group
Low Speed Serial Trigger
Syntax
TRIGger:A:CAN:DATa:SOUrce {CH1|CH2|CH3|CH 4}
TRIGger:A:CAN:DATa:SOUrce?
Arguments
CH<x> specifies the source for the CAN trigger.
Examples
TRIGGER:A:CAN:DATA:SOURCE CH1 sets the CAN trigger data source to
channel CH1.
TRIGGER:A:CAN:DATA:SOURCE? might return
:TRIGGER:A:CAN:DATA:SOURCE CH4, indicating that the CAN trigger data
source is set to channel CH4.
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-553