Commands Listed in Alphabetical Order
Group
Horizontal
Syntax
HORizontal:MAIn:UNIts:STR ing <string>
HORizontal:MAIn:UNIts:STR ing?
Related Commands
HORizontal:MAIn:UNIts
Arguments
<string> is the time base units string.
Examples
HORIZONTAL:MAIN:UNITS:STR ING “LUM” sets the units string for the time
base trigger delay to lumens.
HORIZONTAL:MAIN:UNITS:STR ING? might return
:HORIZONTAL:MAIN:UNITS:ST RING "lum", indicating that the units string for
the time base trigger delay is set to lumens.
HORizontal:MODE
This command set or queries the horizontal mode. Auto mode is the factory
default.
Group
Horizontal
Syntax
HORizontal:MODE {AUTO|CONStant|MANu al}
HORizontal:MODE?
Related Commands
HORizontal:MODE:AUTO:LIMITrecordlen, HORizontal:MODE:
RECOrdlength, HORizontal:MODE:SAMPLERate, HORizontal:MODE:SCAle
Arguments
AUTO selects the automatic horizontal model. Auto mode attempts to keep
record length constant as you change the time per division setting. Record length
is r
ead only.
CONSTANT selects the constant horizontal model. Constant mode attempts to
ke
ep sample rate constant as you change the time per division setting. Record
length is read only.
M
ANUAL selects the manual horizontal model. Manual mode lets you change
sample mode and record length. Time per division or Horizontal scale is read only.
2-260 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual