Commands Listed in Alphabetical Order
ACQuire:NUMS
AMples
This command sets or queries the minimum number of acquired samples that
makeupawaveformdatabase(WfmDB)waveformforsinglesequencemodeand
Mask Pass/Fa
il Completion Test. This is equivalent to setting the Waveform
Database Samples in the Acquisition Mode side menu.
Group
Acquisition
Syntax
ACQuire:NUMSAMples {<NR1>}
ACQuire:NUMSAMples?
Arguments
NR1 is the minimum number of acquired sa
mples that make up a waveform
database (WfmDB) waveform for single sequence mode and Mask Pass/Fail
Completion Test. The default value is 16,000 samples. The range is 5,000 to
2,147,400,000 samples.
In real time mode, if the entire trace is displayed, one triggered acquisition should
acquire 5000 samples for a record length of 5000. If you want 5001 samples, the
instrument will need to make two acquisitions (10,000 samples). If you request
fewer samples than the record length, the number of samples will be the same as
the record length.
Examples
ACQUIRE:NUMSAMPLES 5000 specifies that 5000 samples will be acquired in
WfmDB m ode for single sequence mode and Mask Pa ss/Fail Completion Test.
ACQUIRE:NUMSAMPLES? might return :ACQUIRE:NUMSAMples 10000,
indicating that there are 10,000 samples to be acquired in WfmDB mode for
single sequence and Mask Pass/Fail Completion Test.
ACQuire:SAMPlingmode
This command sets or queries the sampling mode.
Group
Acquisition
Syntax
ACQuire:SAMPlingmode {RT|IT|ET}
ACQuire:SAMPlingmode?
Arguments
RT sets the sampling mode to real time only.
IT sets the sampling mode to interpolation allowed (OFF in REPET).
2-80 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual