Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:METHod
This command specifies or queries the method used to calculate the 0% and
100% reference level.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:METHod {HIStogram|MIN Max|MEAN}
MEASUrement:MEAS<x>:METHod ?
Arguments
HIStogram sets the high and low waveform levels statistically using a histogram
algorithm.
MINMax sets the high and low waveform levels to M AX and MIN, respectively.
MEAN sets the high and low waveform levels to their mean.
Examples
MEASUREMENT:MEAS1:METHOD HISTOGRAM specifies that the high and low
reference levels are set statistically.
MEASUREMENT:MEAS1:METHOD? might return :MEASU REMENT:MEAS1:METHOD
MINMAX
, indicating that the reference levels are set to MIN a nd MAX.
MEASUrement:MEAS<x>:MINImum? (Query Only)
This query-only command returns the minimum value found for this measurement
since t he last statistical reset. Measurements are specified by x, which ranges
from 1 through 8.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINImu m?
Examples
MEASUREMENT:MEAS1:MINIMUM? might return
:MEASUREMENT:MEAS1:MINIMUM 1.75E-09.
MEASUrement:MEAS<x>:NOISe
This command sets or queries whether the noise measurement is made on the high
or low level of the waveform. Sending this command is equivalent to selecting
Ref Levs > Eye > Top Level or Base Level in the Comm tab of the Meas urement
Setup dialog box. The Eye section is displayed only if you have an eye-pattern
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-387