Commands Listed in Alphabetical Order
TRIGger:{A|B
}:PULse:RUNT? (Query Only)
This query-only command returns the current runt trigger p arameters. This
command query is equivalent to selectingRuntSetupfromtheTrigmenuandthen
viewing the c
urrent settings.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:RUNT?
Related C
ommands
TRIGger:{A|B}:PULse:GLItch?, TRIGger:{A|B}:PULse:TIMEOut?,
TRIGger:{A|B}:PULse:TRANsition?, TRIGger:{A|B}:PULse:WIDth?
Example
s
TRIGGER:A:PULSE:RUNT? might return
:TRIGGER:A:PULSE:RUNT:POL ARITY POSITIVE;POLARITY:C H1
POSITIVE;CH2 POSITIVE;CH3 POSITIVE;CH4
POSIT
IVE;:TRIGGER:A:PULSE:RUNT:T HRESHOLD:HIGH 1.2000;LOW
800.0000E-3;:TRIGGER:A:PU LSE:RUNT:WHEN OCCURS;WIDTH
500.0000E-12;QUALIFY OCCURS
.
TRIGger:{A|B}:PULse:RUN
T:POLarity
This command sets or queries the polarity for the A or B pulse runt trigger. This
command is equivalent to selecting Runt Setup from the Trig menu and then
choosing the Polarity se
tting.
Group
Trigger
Syntax
TRIGger:{A|B}:PULse:RUNT: POLarity {POSITIVe|NEGAti ve|EITher}
TRIGger:{A|B}:PULse:RUNT: POLarity?
Related Commands
TRIGger:{A|B}:PULse:RUNT:THReshold?
Arguments
POSITIVe indicates that the rising edge crosses the low threshold and the falling
edge recrosses the low threshold without either edge ever crossing the high
threshold.
NEGAtive indicates that the falling edge crosses the high threshold and the
rising edge recrosses the high threshold without either edge ever crossing the
low threshold.
EITher indicates either negative or positive polarity.
2-526 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual