Commands Listed in Alphabetical Order
Examples
MEASUREMENT:I
MMED:SOURCE1:SIGTYPE PULSE
selects Pulse as the signal
type for source 1.
MEASUREMENT:
IMMED:SOURCE<X>:SIGTYPE?
might return
:MEASUREMENT:IMMED:SOURCE 1:SIGTYPE EYE indicating that the signal
type for source 1 is EYE.
MEASUremen
t:IMMed:TYPe
This command sets or queries the immediate measureme nt type.
Group
Measurement
Syntax
MEASUrement:IMMed:TYPe {AMPlitude|AREa|
BURst|CARea|CMEan|CRMs|DE Lay|DISTDUty|
EXTINCTDB|EXTINCTPCT|EXTI NCTRATIO|
EYEHeight|EYEWIdth|FALL|F REQuency|HIGH|
HITs|LOW|MAXimum|MEAN|MED ian|MINImum|
NCROss|NDUty|NOVershoot|N WIdth|PBASe|
PCROss|PCTCROss|PDUty|PEA KHits|PERIod|
PHAse|PK2Pk|PKPKJitter|PK PKNoise|
POVershoot|PTOP|PWIdth|QF ACtor|RISe|
RMS|RMSJitter|RMSNoise|SI GMA1|SIGMA2|
SIGMA3|SIXSigmajit|SNRati o|STDdev| UNDEFINED|WAVEF ORMS}
MEASUrement:IMMed:TYPe?
Arguments
AMPlitude measures the amplitude of the selected waveform. In other words, it
measures the high value less the low value measured over
the entire waveform or
gated region.
Amplitude = High - Low
AREa measures the voltage over time. The a rea is over the entire waveform or
gated region and is measured in volt-seconds. The ar
ea measured above the
ground is positive, while the area below ground is negative.
BURst measures the duration of a burst. The measurem
entismadeovertheentire
waveform or gated region.
CARea (cycle area) measures the voltage over time.
In other words, it measures, in
volt-seconds, the area over the first cycle in the waveform or the first cycle in the
gated region. The area measured above the common reference point is positive,
while the area below the common reference point is negative.
CMEan (cycle mean) measures the a rithmetic mea n over the first cycle in the
waveform or the first cycle in the gated region.
2-378 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual