Commands Listed in Alphabetical Order
MASK:TESt:LO
G:FAILure
This command sets or returns the log status on pass/fail test failure mode. When
enabled, this command causes the instrument to log the current date and time to a
file when the p
ass/fail status changes to ‘Failing’. The file name is determined
from the current date and time. The user cannot change this name.
Group
Mask
Syntax
MASK:TESt:LOG:FAILure {OFF|ON|<NR1>}
MASK:TESt:LOG:FAILure?
Related Commands
MASK:TESt:SAVEWFM:FILEName
Arguments
ON turns on the pass/fail log o n failure.
OFF turns off the pass/fail log on failure. This is the default.
<NR1> = 0 turns off the pass/fail log on failure, and any other integer turns on the
pass/fail log on failure.
Examples
MASK:TEST:LOG:FAILURE OFF turns off the pass/fail log on failure.
MASK:TEST:LOG:FAILURE? might return :MASK:TE ST:LOG:FAILURE ON,
indicating that the pass/fail log on failure is turned on.
MASK:TESt:REPeat
This command sets or returns the repeat pass/fail test on completion mode. When
enabled, this command repeats the pass/fail test cycle at the completion o f the
previous test cycle.
Group
Mask
Syntax
MASK:TESt:REPeat {OFF|ON|<NR1>}
MASK:TESt:REPeat?
Related Commands
MASK:COUNt:VIOLATIONS?
Arguments
ON turns on repeat on completion.
OFF turns off repeat on completion.
2-328 DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual