Commands Listed in Alphabetical Order
Examples
CH1:PROBE:INP
UTMODE:CMOFFSET 5.0
sets the requested common mode
offset control of the probe that is attached to the Channel 1 to its 5 Ampere range.
CH1:PROBE:INPUTMODE:CMOFFS ET? might return
:CH1:PROBE:INPUTMODE:CMOFF SET 0.0000, indicating that the requested
common mode offset control of the probe that is attached to the Channel 1 is
set to 0.0.
CH<x>:PRObe:INPUTMode:DMOFFSet
This command sets or queries the requested common mode differential mode
offset control of the probe that is attached to the specified channel.
Group
Vertical
Syntax
CH<x>:PRObe:INPUTMode:DMOF FSet
CH<x>:PRObe:INPUTMode:DMOF FSet?
Arguments
NR3 specifies the differential mode offset control.
Examples
CH1:PROBE:INPUTMODE:DMOFFS ET 5.0 sets the requested differential mode
offset control of the probe that is attached to the Channel 1 to its 5 Ampere range.
CH1:PROBE:INPUTMODE:DMOFFS ET? might return
:CH1:PROBE:INPUTMODE:DMOFF SET 0.0000, indicating that the requested
differential mode offset control of the probe that is attached to the Channel
1 is set to 0.0.
CH<x>:PRObe:RANge
This command controls or queries the attenuation range of the probe on the
designated channel.
Group
Vertical
Syntax
CH<x>:PRObe:RANge
{ATTEN1X|ATTEN1_25X|ATTEN1 _5X|ATTEN1_75X|ATTEN2X|
ATTEN2_5X|ATTEN3X|ATTEN3_5 X|ATTEN4X|ATTEN4_5X|
ATTEN5X|ATTEN5_5X|ATTEN6X| ATTEN6_5X|ATTEN7X|
ATTEN7_5X|ATTEN8X|ATTEN8_5 X|ATTEN9X|ATTEN9_5X|
ATTEN10X|ATTEN12_5X|ATTEN1 5X|ATTEN17_5X|ATTEN20X|
ATTEN25X|ATTEN30X|ATTEN35X |ATTEN40X|ATTEN45X|
ATTEN50X|ATTEN55X|ATTEN60X |ATTEN65X|ATTEN70X|
DPO7000, DPO70000/B and DSA7000/B Series Programmer Manual 2-121