EasyManua.ls Logo

Anritsu MS9740A - Calculate:data:apr:mpkc; Calculate:data:apr:wmpk

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.4 Device Message Details
3-61
3
Message Details
:CALCulate:DATA:APR:MPKC?
Function
This command queries the number of the detected multi peaks.
Syntax
:CALCulate:DATA:APR:MPKC?
Response Data
<numeric>
<numeric>: Multi peak count
Example of Use
:CALCulate:DATA:APR:MPKC?
>1
:CALCulate:DATA:APR:WMPK?
Function
This command queries the analysis results at the MultiPeak display of
the WDM application function for the specified peak No.
Syntax
:CALCulate:DATA:APR:WMPK? <numeric>
<numeric>: No. of peak to query analysis results
Response Data
<numeric>,<numeric>
No.
Parameter type
Description
1
<numeric>
Peak wavelength of specified peak (m)
2
<numeric>
Peak level of specified peak (dBm)
Example of Use
:CALCulate:DATA:APR:WMPK? 10
>+1.55276000E-006,-1.90200000E+000

Table of Contents