3.4 Device Message Details
3-193
3
:DISPlay:WINDow:SYSTem
Function
This command switches the Config screen and the Spectrum
Measurement screen. It can be switched using a system management
command or a measurement command.
This command queries the types of commands that can be used and the
screen display.
For the system management and measurement commands, refer to 3.3.2,
“System Management and Measurement Commands”.
Syntax
:DISPlay:WINDow:SYSTem CONFIG|OSA[,ACT|INACT|MIN]
:DISPlay:WINDow:SYSTem? CONFIG|OSA
CONFIG: Activates Config screen display and system management
command
OSA: Activates measurement screen display and measurement
command
ACT: Displays front-most screen and permits operation (active
status)
When the Config screen is set to Active, the system
management \command is available.
When the measurement screen is set to Active, the
measurement command is available.
INACT: Makes screen operations inactive
MIN: Minimizes screen display size
ACT is assumed if the setting is omitted.
Response Data
CRRENT|IDLE|RUN|UNLOAD,ACT|INACT|MIN|NON
Executes and makes operation target
Loads but does not execute
Executes, but does not make operation target
Displays not active screen
Displays minimized screen
Example of Use
:DISPlay:WINDow:SYSTem OSA,MIN
:DISPlay:WINDow:SYSTem? OSA