EasyManua.ls Logo

Anritsu MS9740A - Calculate:parameter[:Category]:Wdm:falgo

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-154
:CALCulate:PARameter[:CATegory]:WDM:FALGo
Function
This command sets and queries the Fitting Curve in Noise Parameter for
the WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:FALGo <switch>
:CALCulate:PARameter[:CATegory]:WDM:FALGo?
<switch>
Fitting Curve
LINear
Linear
GAUSs
Gauss
3RD
3rd POLY
4TH
4th POLY
5TH
5th POLY
Response Data
0|1|3|4|5
0: Linear
1: Gauss
3: 3rd POLY
4: 4th POLY
5: 5th POLY
Example of Use
:CALCulate:PARameter:CATegory:WDM:FALGo 3RD
:CALCulate:PARameter:CATegory:WDM:FALGo?
>3

Table of Contents