3.4 Device Message Details
3-111
3
:CALCulate:PARameter[:CATegory]:DFBLd:SRES
Function
This command sets and queries the level resolution to detect the side
mode in DFB-LD application.
Syntax
:CALCulate:PARameter[:CATegory]:DFBLd:SRES <numeric>[DB]
:CALCulate:PARameter[:CATegory]:DFBLd:SRES?
<numeric>: Level resolution
Range 0.10 to 10.00 dB
Resolution 0.01 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Level resolution (dB)
+1.00000000–E001 to +1.00000000+E001
Example of Use
:CALCulate:PARameter:CATegory:DFBLd:SRES 2
:CALCulate:PARameter:CATegory:DFBLd:SRES?
>+2.00000000+E000