Chapter 3 Message Details
3-104
:CALCulate:PARameter[:CATegory]:AMP:MSELect
Function
This command selects and queries the saving destination of the
measurement data at the Optical AMP application.
Note:
PASE can be specified as the measured data save destination of
the measured data when an optical amplifier measurement method
is polarization nulling (PLZN Nulling). If another measurement
method is set, an error is returned when PASE is specified.
Syntax
:CALCulate:PARameter[:CATegory]:AMP:MSELect <switch>
:CALCulate:PARameter[:CATegory]:AMP:MSELect?
<switch>: Saving destination of measurement data
{PIN|POUT|PASE}
Response Data
PIN|POUT|PASE
Example of Use
:CALCulate:PARameter:CATegory:AMP:MSELect POUT
:CALCulate:PARameter:CATegory:AMP:MSELect?
>POUT