Chapter 3 Message Details
3-160
:CALCulate:PARameter[:CATegory]:WDM:NBW
Function
This command sets and queries the Noise BW value for Normalization
performed by the WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:NBW
<numeric>[<unit_m>]
:CALCulate:PARameter[:CATegory]:WDM:NBW?
<numeric>: Noise BW setting value
Range 0.1 to 1.0 nm
Resolution 0.1 nm
If a suffix is omitted, the unit will be m.
Response Data
<numeric>
<numeric>: Noise BW setting value (m)
+1.00000000E–010 to +1.00000000E–009
When this command is executed, Normalization for Noise Parameter is
set to On.
Example of Use
:CALCulate:PARameter:CATegory:WDM:NBW 0.5NM
:CALCulate:PARameter:CATegory:WDM:NBW?
>+5.00000000E-010
:CALCulate:PARameter[:CATegory]:WDM:NNORmalize
Function
This command sets and queries the Normalization and Noise BW for the
WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:NNORmalize
{0|1|OFF|ON}[,<numeric>[<unit_m>]]
:CALCulate:PARameter[:CATegory]:WDM:NNORmalize?
0|OFF: Sets the Normalization to Off.
1|ON: Sets the Normalization to On.
<numeric>: Noise BW setting value