EasyManua.ls Logo

Anritsu MS9740A - Calculate:parameter[:Category]:Led

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-128
:CALCulate:PARameter[:CATegory]:LED
Function
This command exercutes the LED application and specifies the
parameters.
This command queries the parameters for the LED application.
Syntax
:CALCulate:PARameter[:CATegory]:LED
<numeric>,<numeric>,<numeric>
:CALCulate:PARameter[:CATegory]:LED?
No.
Parameter Type
Range
Description
1
<numeric>
0.1 to 50.00
dB
Res. 0.01 dB
Cut Level
2
<numeric>
10.00 to
10.00dB
Res. 0.01 dB
Total power correction value
(dB)
3
<numeric>
1.00 to 10.00
Res. 0.01
K: Standard deviation factor
Response Data
<numeric>,<numeric>,<numeric>
No.
Parameter Type
Description
1
<numeric>
Cut Level (dB)
2
<numeric>
Total power correction value
(dB)
3
<numeric>
K: Standard deviation factor
Example of Use
:CALCulate:PARameter:CATegory:LED 35,0,2.35
:CALCulate:PARameter:CATegory:LED?
>7,+3.5000000E+001,+0.0000000E+000,+2.3500000E+000

Table of Contents