3.4 Device Message Details
3-219
3
Message Details
:MMEMory:LOAD:ATRace
Function
This command reads the parameters and data for 10 traces from the
XML file saved in the specified device.
If the specified device or file cannot be found and an error is generated,
the standard event status execution error bit becomes 1.
The XML files in the following folder of the specified device can be read.
x:\Anritsu Corporation\Optical Spectrum Analyzer\User Data\All Trace Data
Syntax
:MMEMory:LOAD:ATRace <file_name>[,<switch>]
It is not necessary to add the extension to file_name.
<switch>: Specify a device from the following.
INTernal: Internal storage media (DriveD)
EXTernal: USB storage media
When there are two or more USB storage media, the drive
having the first drive letter (alphabetically) is specified
automatically.
If <switch> is omitted, it defaults to INTernal.
Example of Use
:MMEMory:LOAD:ATRace 'LED_125M(025)',EXTernal