3.4 Device Message Details
3-209
3
Message Details
To check that the sweeping is completed, use the *OPC? command.
When the sweeping has completed, the response to the *OPC? command
is 1.
:INITiate:IMMediate:SEQuence[1|2]
Function
This command starts the Pin or Pout measurement by the Optical AMP
(WDM) application.
Syntax
:INITiate:IMMediate:SEQuence[1|2]
1: Pin
2: Pout
If [1|2] is omitted, Pin is measured.
:INITiate:POWer:DATA?
Function
This command sets and queries the power monitor wavelength.
Syntax
:INITiate:POWer:DATA?
Response Data
<numeric>
<numeric>: Measurement results of the power monitor (dBm)
When sending :INITiate:POWer:DATA? without setting to power monitor,
***
is queried as response data.
Example of Use
:INITiate:POWer:DATA?
>-4.88000000E+000
:INITiate:POWer:OFF
Function
This command exits the power monitor measurement.
Syntax
:INITiate:POWer:OFF