EasyManua.ls Logo

Anritsu MS9740A - System:information; System:preset

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-250
:SYSTem:INFormation?
Function
This command queries the system information.
This message is a system management command.
Syntax
:SYSTem:INFormation? ALL|MODEL|SERIAL|TYPE
ALL: Queries product name, model name, and serial number.
MODEL: Queries model name.
SERIAL: Queries serial number.
TYPE: Queries product name.
Response Data
<string>|<numeric>|{<string>,<string>,<numeric>}
<string>: Character displaying product name model
<numeric>: Integer value indicating serial number
Example of Use
:SYSTem:INFormation? ALL
>Optical Spectrum Analyzer,MS9740A,610000001
:SYSTem:INFormation? MODEL
>MS9740A
:SYSTem:INFormation? SERIAL
>610000001
:SYSTem:INFormation? TYPE
>Optical Spectrum Analyzer
:SYSTem:PRESet
Function
This command initializes the measurement parameter.
As for the initialized parameters and default values, refer to Appendix B,
Initial Valuesin the
MS9740A Optical Spectrum Analyzer Operation
Manual
.
Syntax
:SYSTem:PRESet

Table of Contents