Chapter 3 Message Details
3-250
:SYSTem:INFormation?
Function
This command queries the system information.
This message is a system management command.
Syntax
:SYSTem:INFormation? ALL|MODEL|SERIAL|TYPE
ALL: Queries product name, model name, and serial number.
MODEL: Queries model name.
SERIAL: Queries serial number.
TYPE: Queries product name.
Response Data
<string>|<numeric>|{<string>,<string>,<numeric>}
<string>: Character displaying product name model
<numeric>: Integer value indicating serial number
Example of Use
:SYSTem:INFormation? ALL
>Optical Spectrum Analyzer,MS9740A,610000001
:SYSTem:INFormation? MODEL
>MS9740A
:SYSTem:INFormation? SERIAL
>610000001
:SYSTem:INFormation? TYPE
>Optical Spectrum Analyzer
:SYSTem:PRESet
Function
This command initializes the measurement parameter.
As for the initialized parameters and default values, refer to Appendix B,
“Initial Values” in the
MS9740A Optical Spectrum Analyzer Operation
Manual
.
Syntax
:SYSTem:PRESet