3.4 Device Message Details
3-105
3
:CALCulate:PARameter[:CATegory]:AMP:PARameter
Function
This command sets and queries the measurement parameter in the
Optical AMP application.
Syntax
:CALCulate:PARameter[:CATegory]:AMP:PARameter
<switch>,<switch>,<switch>,<numeric>,…,<numeric>
:CALCulate:PARameter[:CATegory]:AMP:PARameter?
0: Spect Div Off: Spectrum division off
1: Spect Div On: Spectrum division on
2: PLZN Nulling: Polarization nulling
3: Pulse Method: Pulse method
4: WDM Measure: WDM measurement
0: Gauss Fitting ASE Level found by Gauss method
1: Mean Fitting ASE Level found by averaging value
0.10 to 100.00 nm
Res. 0.01 nm
Fitting Span (m)
Wavelength range for calculating ASE level
Res. 0.01 nm
Wavelength range excluded from ASE level
calculation
Set a small value than Fitting Span.
–10.00 to 10.00
Res. 0.01 dB
Pin Loss
Optical signal level loss correction coefficient
–10.00 to 10.00
Res. 0.01 dB
Pout Loss
Optical signal level loss correction coefficient
0.100 to 10.000
Res. 0.001
NF Calibration
Noise figure correction coefficient
0.00 to 30.00
Res. 0.01 dB
O.BPF Level Calibration
Optical filter loss correction coefficient
0.00 to 999.99 nm
Res. 0.01 nm
O.BPF Band Width
Optical filter passsband width
–10.00 to 10.00
Res. 0.01 dB
Pol Loss
Polarization controller loss correction coefficient
Note:
Parameters 5
th
to 11
th
are common parameters at optical amplifier
measurement depending on the second <switch> (measurement
method) setting.
Depending on the measurement method, the 5
th
to 11
th
<numeric>
is an unnecessary parameter but it cannot be omitted. In this case,
set any in-range value at the 5
th
to 11
th
<numeric>.