3.4 Device Message Details
3-159
3
:CALCulate:PARameter[:CATegory]:WDM:NAUSer
Function
This command sets and queries the User Specify Area in Noise
Parameter for the WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:NAUSer
<numeric>[<unit_m>],<numeric>[<unit_m>],<numeric>[<unit_
m>],<numeric>[<unit_m>]
:CALCulate:PARameter[:CATegory]:WDM:NAUSer?
Resolution 0.01 nm
If a suffix is omitted, the unit will be m.
Response Data
<numeric>,<numeric>,<numeric>,<numeric>
<numeric>: Wavelength position/interval (m)
+1.00000000E–011 to +1.00000000E–007
Example of Use
:CALCulate:PARameter:CATegory:WDM:NAUSer
50.00NM,10.00NM,60.00NM,15.00NM
:CALCulate:PARameter:CATegory:WDM:NAUSer?
>+5.00000000E-008,+1.00000000E-008,+6.00000000E-008,+1.5
0000000E-008