3.4 Device Message Details
3-189
3
2: Terminates wavelength calibration due to lack of optical level
3: Terminates wavelength calibration due to other abnormal
phenomena
Example of Use
:CALibration:WAVelength:INTernal:IMMediate
:CALibration:WAVelength:INTernal:IMMediate?
>1
:CALibration:WAVelength:INTernal[:NORMal]
Function
This command performs wavelength calibration when using a reference
light source option and creates the wavelength calibration data.
This command queries the wavelength calibration execution result.
When wavelength calibration is completed, 1 is written to bit 4
(execution completion bit) of the end event status register.
Syntax
:CALibration:WAVelength:INTernal[:NORMal]
:CALibration:WAVelength:INTernal[:NORMal]?
Response Data
0|1|2|3
0: Ends wavelength calibration
1: Wavelength calibration in progress
2: Terminates wavelength calibration due to lack of optical level
3: Terminates wavelength calibration due to other abnormal
phenomena
Example of Use
:CALibration:WAVelength:INTernal:NORMal
:CALibration:WAVelength:INTernal:NORMal?
>1