3.4 Device Message Details
3-109
3
Message Details
:CALCulate:PARameter[:CATegory]:AMP:RESCalibrate
Function
This command calibrates the resolution of the optical spectrum analyzer
for the Optical AMP application.
Bit 4 (execution completion bit) of the end event status register (ESR2) is
set to 1 after the completion of resolution calibration.
This command queries the status of the resolution calibration in the
Optical AMP application.
Syntax
:CALCulate:PARameter[:CATegory]:AMP:RESCalibrate 0|1
:CALCulate:PARameter[:CATegory]:AMP:RESCalibrate?
0: Initializes current resolution calibration data
1: Executes resolution calibration
Response Data
0|1|2|3
0: Uses default resolution calibration value
1: Resolution calibration finished normally
2: Calibrating resolution
3: Resolution calibration finished abnormally
Example of Use
:CALCulate:PARameter:CATegory:AMP:RESCalibrate 1
:CALCulate:PARameter:CATegory:AMP:RESCalibrate?
>2
:CALCulate:PARameter:CATegory:AMP:RESCalibrate?
>3