Chapter 3 Message Details
3-174
Response Data
<numeric>
<numeric>: Slice Level (dB)
+1.00000000E
–
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WFIL:SLICe 32
:CALCulate:PARameter:CATegory:WFIL:SLICe?
>+3.20000000E+001
:CALCulate:PARameter[:CATegory]:WFIL:TCL
Function
This command sets and queries the cut level to measure the wavelength
at WDM filter application.
Syntax
:CALCulate:PARameter[:CATegory]:WFIL:TCL
<numeric>[DB],<numeric>[DB]
:CALCulate:PARameter[:CATegory]:WFIL:TCL?
Parameter 1: Cut Level A
Parameter 2: Cut Level B
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>,<numeric>
Parameter 1: Cut Level A (dB)
Parameter 2: Cut Level B (dB)
<numeric>: Cut Level (dB)
+1.00000000E
–
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WFIL:TCL 6,36
:CALCulate:PARameter:CATegory:WFIL:TCL?
>+6.00000000E+000,+3.60000000E+001