Anritsu MS9740A - Calculate:parameter[:Category]:Wfil:tcl

Anritsu MS9740A
320 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Message Details
3-174
Response Data
<numeric>
<numeric>: Slice Level (dB)
+1.00000000E
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WFIL:SLICe 32
:CALCulate:PARameter:CATegory:WFIL:SLICe?
>+3.20000000E+001
:CALCulate:PARameter[:CATegory]:WFIL:TCL
Function
This command sets and queries the cut level to measure the wavelength
at WDM filter application.
Syntax
:CALCulate:PARameter[:CATegory]:WFIL:TCL
<numeric>[DB],<numeric>[DB]
:CALCulate:PARameter[:CATegory]:WFIL:TCL?
Parameter 1: Cut Level A
Parameter 2: Cut Level B
Range 0.1 to 50.0 dB
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>,<numeric>
Parameter 1: Cut Level A (dB)
Parameter 2: Cut Level B (dB)
<numeric>: Cut Level (dB)
+1.00000000E
001 to +5.00000000E+001
Example of Use
:CALCulate:PARameter:CATegory:WFIL:TCL 6,36
:CALCulate:PARameter:CATegory:WFIL:TCL?
>+6.00000000E+000,+3.60000000E+001

Table of Contents