3.4 Device Message Details
3-201
3
Message Details
>1
:DISPlay[:WINDow[1]]:TRACe:Y[1|2][:SCALe]:PDIVision
Function
This command sets the level scale to Log and scale division (dB/div)
This command queries the Log scale.
Syntax
:DISPlay[:WINDow[1]]:TRACe:Y[1|2][:SCALe]:PDIVision
<numeric>[DB]
:DISPlay[:WINDow[1]]:TRACe:Y[1|2][:SCALe]:PDIVision?
The process will not change regardless of whether a number [1|2] is
specified or omitted.
Range 0.1 to 10.0 dB (Absolute value display)
0.1 to 20.0 dB (Relative value display)
Resolution 0.1 dB
If a suffix is omitted, the unit will be dB.
Response Data
<numeric>
<numeric>: Log scale value (dB)
+1.00000000E
001 to +1.00000000E+001 (Absolute value
display)
+1.00000000E
001 to +2.00000000E+001 (Relative value
display)
Example of Use
:DISPlay:WINDow:TRACe:Y:SCALe:PDIVision 0.2
:DISPlay:WINDow:TRACe:Y:SCALe:PDIVision?
>+2.00000000E-001

Table of Contents

Related product manuals