:CALibration:WAVelength[:EXTernal][:NORMal]
Function
This command performs wavelength calibration when using an external
light source and creates the wavelength calibration data.
This command queries the wavelength calibration execution result.
When wavelength calibration is completed, 1 is written to bit 4
(execution completion bit) of the end event status register.
Syntax
:CALibration:WAVelength[:EXTernal][:NORMal]
:CALibration:WAVelength[:EXTernal][:NORMal]?
Response Data
0|1|2|3
0: Ends wavelength calibration
1: Wavelength calibration in progress
2: Terminates wavelength calibration due to lack of optical level
3: Terminates wavelength calibration due to other abnormal
phenomena
Example of Use
:CALibration:WAVelength:EXTernal:NORMal
:CALibration:WAVelength:EXTernal:NORMal?
>3
:CALibration:WAVelength:INTernal[:IMMediate]
Function
This command performs wavelength calibration when using a reference
light source option and creates the wavelength calibration data.
This command queries the wavelength calibration execution result.
When wavelength calibration is completed, 1 is written to bit 4
(execution completion bit) of the end event status register.
Syntax
:CALibration:WAVelength:INTernal[:IMMediate]
:CALibration:WAVelength:INTernal[:IMMediate]?
Response Data
0|1|2|3
0: Ends wavelength calibration
1: Wavelength calibration in progress