3.4 Device Message Details
3-163
3
:CALCulate:PARameter[:CATegory]:WDM:POINt
Function
This command sets and queries Point and Noise Position in Noise
Parameter for the WDM application.
Syntax
:CALCulate:PARameter[:CATegory]:WDM:POINt
<switch>,{<numeric>|OFF}
:CALCulate:PARameter[:CATegory]:WDM:POINt?
Parameter 1: Position of noise level detection
Parameter 2:
<numeric>: Noise Position
Range 0.01 to 20.00 nm
Resolution 0.01 nm
If a suffix is omitted, the unit will be m.
OFF: Sets the Noise Position measurement to Off.
Response Data
<switch>,{<numeric>|0}
Parameter 1: Position of noise level detection
Parameter 2:
<numeric>: Noise Position
+1.00000000E–011 to +2.00000000E–008
0 Off
Example of Use
:CALCulate:PARameter:CATegory:WDM:POINt AVERAGE,OFF
:CALCulate:PARameter:CATegory:WDM:POINt?
>AVERAGE,0