3.4 Device Message Details
3-131
3
Message Details
:CALCulate:PARameter[:CATegory]:NF:AREA
Function
This command sets and queries the ASE Area Parameter for the Optical
AMP (WDM) application.
Syntax
:CALCulate:PARameter[:CATegory]:NF:AREA
CENTER|<numeric>[<unit_m>],<numeric>[<unit_m>]
:CALCulate:PARameter[:CATegory]:NF:AREA?
No.
Parameter type
Range
Description
1
CENTER
<numeric>
0.10 to 100.00 nm
Sets the halfway point
between channels as
the interpolation range.
Fitting Span
2
<numeric>
0.10 to 100.00 nm
Masked Span
Resolution 0.01 nm
If a suffix is omitted, the unit will be m.
Response Data
CENTER|<numeric_value>,<numeric>
No.
Parameter type
Description
1
CENTER
<numeric>
Sets the halfway point between
channels as the interpolation range.
Fitting Span (m)
2
<numeric>
Masked Span (m)
<numeric>: Fitting Span/ Masked Span (m)
+1.00000000E
010 to +1.00000000E007
Example of Use
:CALCulate:PARameter:CATegory:NF:AREA 1.00e-9,0.50e-9
:CALCulate:PARameter:CATegory:NF:AREA?
>+1.00000000E-009,+5.00000000E-010