3.4 Device Message Details
3-185
3
:CALibration:ALIGnment
Function
This command
executes optical alignment. When alignment is complete, bit
4 of the end event status register (execution complete bit) is set to 1.
Syntax
:CALibration:ALIGnment
:CALibration:ALIGnment?
Response Data
0|1|2|3
0: Normal end
1: During alignment
2: Aborted optical alignment due to lack of optical level
3: Aborted optical alignment due to other abnormality
Example of Use
:CALibration:ALIGnment
:CALibration:ALIGnment?
>1
:CALibration:ALIGnment:PRESet
Function
This command returns the optical alignment data to the factory defaults.
Syntax
:CALibration:ALIGnment:PRESet
:CALibration:AUToalign:ENABle
Function
This command sets and queries whether performing optical alignment
using the optional reference light source
.
Syntax
:CALibration:AUToalign:ENABle 0|1|OFF|ON
:CALibration:AUToalign:ENABle?
0|OFF: Disables performing the optical alignment.
1|ON: Enables performing the optical alignment.
Response Data
0|1