Chapter 3 Message Details
3-40
:CALCulate[1|2|3|4|5|6]:CENTermass[:DATA]?
Function
This command queries the Mean Wl (FWHM) value measured by the
LED application.
Syntax
:CALCulate[1|2|3|4|5|6]:CENTermass[:DATA]?
Specify the Active trace (A to F) by specifying a number from 1 to 6.
When omitted, the Active trace is not changed.
Response Data
<numeric>
<numeric>: Mean Wl (FWHM) measured value (m)
Example of Use
:CALCulate4:CENTermass:DATA?
>+1.31000000E-006
:CALCulate[1|2|3|4|5|6]:CENTermass:STATe
Function
This command sets and queries the On/Off state of the LED application
display.
Syntax
CALCulate[1|2|3|4|5|6]:CENTermass:STATe 0|1|OFF|ON
CALCulate[1|2|3|4|5|6]:CENTermass:STATe?
Specify the Active trace (A to F) by specifying a number from 1 to 6.
When omitted, the Active trace is not changed.
0|OFF: LED application not displayed
1|ON: LED application displayed
Response Data
0|1
Example of Use
:CALCulate2:CENTermass:STATe ON
:CALCulate2:CENTermass:STATe?
>1