Chapter 3 Message Details
3-210
:INITiate:POWer:WAVelength
Function
This command sets and queries the power monitor wavelength.
When processing is completed, bit 3 of the end event status register is set
to 1.
Syntax
:INITiate:POWer:WAVelength 632.8|850|1300|1550
:INITiate:POWer:WAVelength?
632.8: 632.8 nm
850: 850 nm
1310: 1310 nm
1550: 1550 nm
Response Data
632.8|850|1300|1550
Other functions cannot be executed while the power monitor
measurement is in progress.
To execute the other function, use the
:INITiate:POWer:OFF
command
to terminate the power monitor measurement.
:INITiate:SMODe
Function
This command sets and queries the sweep mode.
Syntax
:INITiate:SMODe <switch>
:INITiate:SMODe?
<switch>: Sweep mode
1|SINGle: Single sweeping
2|REPeat: Repeat sweeping
3|AUTO: Auto Measure
Response Data
1|2|3
To s tar t sweeping, send :INITiate[:IMMediate] after this message.